The Bear Cub 750vs infant ventilator is designed to accommodate a range of patients from the small neonate to the 30kg pediatric. The Bear Cub 750vs and 750psv Infant Ventilators are designed to provide ventilator support for the critical care management of patients ranging from the small neonate (500 grams or larger) to the small pediatric patient (up to 30 kg) with compromised lung function. The 750vs has integrated synchronized ventilation and tidal volume monitoring. It also includes independent base & inspiratory flows and an internal battery back-up.
Bear Cub 750 PSV Ventilator
Warranty | 3-months |
---|---|
Condition | Refurbished |
Manufacturer | Viasys |
Model Number | Viasys PSV Ventilator |
Controls
- Mode: AC, Flow cycle AC, SIMV/IMV, Flow Cycled SIMV, SIMV/PSV, PSV, CPAP
- Ventilator Rate: 1.0 to 150 BPM
- Inspiratory Time: 0.1 to 3.0 sec.
- Inspiratory Flow: 1.0 to 30 LPM
- Base Flow: 1.0 to 30 LPM
- Volume Limit: 5.0 to 300 ml
- PEEP: 0 to 30 cmH2O
- Inspiratory Pressure: 0 to 72 cmH2O
- Manual Breath: x One
- Assist Sensitivity: 0.2 to 5.0 LPM; Minimum to Maximum bar
- Over Pressure Relief: 15 to 75 cmH2O
- % O2: 21 to 100%
- Alarm Loudness: Min 60 to Max. 75 dbA
- Audible Off Volume Limit: LED On/Off
- High Breath Rate: 3.0 to 255 BPM
- Low PEEP/CPAP: -5.0 to 30 cmH2O
- Low Inspiratory Pressure (vs): 1.0 to 65 cmH2O
- Low Inspiratory Pressure (psv): 5 to 41 cmH2O (derived from High Pressure Limit and Low PEEP/CPAP Alarm Settings
- Patient Circuit: LED On/Off
- Failed to Cycle: LED On/Off
- Low Gas Supply: LED On/Off
- Apnea Alarm: 5, 10, 20, 30 sec
- Settings Incompatible: LED On/Off
- Pressure Settings Incompatible: LED On/Off
- Prolonged Inspiratory Pressure: LED On/Off
- Flow Sensor: LED On/Off
- Low Battery: LED On/Off
- Alarm Silence: 60 sec
- Visual Reset: Push Button
- High Pressure Limit: 10 to 75 cmH2O
- Low Minute Volume: Off to 99 lpm
- PEEP: 0 to 30 cmH2O
- Breath Rate: 0 to 255 BPM
- Breath Type (Patient Initiated): LED
- Minute Volume: 0 to 30.0 LPM
- Tidal Volume (Exhaled): 0 to 500 ml
- % Tube Leak: 0 to 100%
- Inspiratory Time: 0.1 to 3.10 sec Expiratory Time 9 to 99.9 sec
- I:E Ratio: 9.9:1 to 1:9.9
- Peak Inspiratory Pressure: 0 to 99 cmH2O
- Mean Airway Pressure: 0 to 75 cmH2O
- Air Pressure: 0 to 100 psig; 0 to 99 psig
- O2 Pressure: 0 to 100 psig; 0 to 99 psig
- Proximal Airway Pressure: -10 to 100 cmH2O
- Hour meter: 0 to 99,999 hours
- Test: Push Button
- Line Power: Green/Red LED
- Battery: LED On/Off
- Inspired Tidal Volume: 0 to 500 ml
- Minimum Expiratory Time: 0.15 seconds
- Maximum Inspiratory Time: 3:10 seconds
- Maximum settable I:E Ratio: 4:1
TOLERANCES:
Monitors- Breath Rate: ± 1 BPM or ± 20 msec, whichever is greater
- Inspiratory Time: ± 0.02 sec
- Expiratory Time: ± 0.02 sec
- I:E Ratio: ± 0.1 or ± 20 msec (on the calculation based on the monitored inspiratory and expiratory times), whichever is greater
- Airway Pressure (Continuous): ± 1 cmH2O (-10 to 20 cmH2O) ± 2 cmH2O (20 to 65 cmH2O) ± 3 cmH2O (65 to 100 cmH2O)
- Peak Inspiratory Pressure: ± 2 cmH2O or ± 5%, whichever is greater
- Mean Airway Pressure: ± 2 cmH2O or ± 3%, whichever is greater
- PEEP: ± 2 cmH2O or ± 5%, whichever is greater
- Air/O2 Pressure: ± 5 psig
- Minute Volume Accuracy: Based on the Tidal Volume and Breath Rate Monitors
- Inspiratory Tidal Volume: ± 1 ml or ± 10%, whichever is greater
- Expiratory Tidal Volume: ± 1 ml or ± 10%, whichever is greater
- % Leak: ± 1 ml or ± 2%, whichever is greater (when compared to the Inspired and Expired Tidal Volume Monitors)
- Hour Meter: ± 2% of reading
- Low PEEP/CPAP: ± 2 cmH2O
- High Breath Rate: ± 1 BPM
- Low Minute Volume: ± 10 ml/min (0 to 99 ml/min) ± 0.1 L/min (0.1 to 9.9 L/min)
- High Pressure Limit: ± 4 cmH2O
- Prolonged Inspiratory Pressure: ± 2 cmH2O
- Volume Limit: ± 2 ml or ± 10% of setting, whichever is greater
- Apnea: ± 1 sec
- Maximum (Pop-off) Pressure: ± 4 cmH2O
- Ventilator Rate: ± 1 BPM or ± 20 msec (applied to the breath interval), whichever is greater
- Inspiratory Rate: ± 0.020 sec (0.10 to 0.50 sec) ± 0.025 sec (0.50 to 3.00 sec)
- Assist Sensitivity: Not a calibrated scale
- PEEP/CPAP and Inspiratory: ± 1 cmH2O or ± 5%, whichever is greater (breath to breath)
- Base/Inspiratory Flow: ± 0.5 LPM or ± 10% of setting, whichever is greater (0 to 60 cmH2O)
- Oxygen %: ± 3%
- Degree of Safety in Presence of Flammable Anesthetics: None
- Mode of Operation: Continuous
Ask A Question
Related Products
-
-
Ventilators
Resmed AirSense 10 AutoSet CPAP with Humidifier
$510.00 View DetailsCondition: Refurbished -
-
-
-
-
-
-
Ventilators
Carefusion LTV1200 Ventilator Teaching Unit
$5,400.00 View DetailsCondition: Refurbished -
Ventilators
Carefusion (Pulmonetics) LTV950 Ventilator – Teaching Unit
$1,485.00 View DetailsCondition: Refurbished -
Ventilators
Carefusion (Pulmonetics) LTV1200 Ventilator
$6,750.00 View DetailsCondition: Refurbished -
Ventilators
Carefusion (Pulmonetics) LTV1000 Ventilator
$3,037.00 View DetailsCondition: Refurbished